Comparison of serum testosterone and estradiol measurements in 3174 European men using platform immunoassay and mass spectrometry; relevance for the diagnostics in aging men

Author:

Huhtaniemi Ilpo T,Tajar Abdelouahid,Lee David M,O'Neill Terence W,Finn Joseph D,Bartfai György,Boonen Steven,Casanueva Felipe F,Giwercman Aleksander,Han Thang S,Kula Krzysztof,Labrie Fernand,Lean Michael E J,Pendleton Neil,Punab Margus,Silman Alan J,Vanderschueren Dirk,Forti Gianni,Wu Frederick C W,_ _

Abstract

BackgroundThe limitations of serum testosterone and estradiol (E2) measurements using non-extraction platform immunoassays (IAs) are widely recognized. Switching to more specific mass spectrometry (MS)-based methods has been advocated, but directly comparative data on the two methods are scarce.MethodsWe compared serum testosterone and E2 measurements in a large sample of middle-aged/elderly men using a common platform IA and a gas chromatography (GC)–MS method, in order to assess their limitations and advantages, and to diagnose male hypogonadism. Of subjects from the European Male Aging Study (n=3174; age 40–79 years), peripheral serum testosterone and E2 were analyzed using established commercial platform IAs (Roche Diagnostics E170) and in-house GC–MS methods.ResultsOver a broad concentration range, serum testosterone concentration measured by IA and MS showed high correlation (R=0.93, P<0.001), which was less robust in the hypogonadal range (<11 nmol/l; R=0.72, P<0.001). The IA/MS correlation was weaker in E2 measurements (R=0.32, P<0.001, at E2 <40.8 pmol/l, and R=0.74, P<0.001, at E2 >40.8 pmol/l). Using MS as the comparator method, IA ascertained low testosterone compatible with hypogonadism (<11 nmol/l), with 75% sensitivity and 96.3% specificity. The same parameters with IA for the detection of low E2 (<40.7 pmol/l) were 13.3 and 99.3%, and for high E2 (>120 pmol/l) 88.4 and 88.6%.ConclusionA validated platform IA is sufficient to detect subnormal testosterone concentrations in the diagnosis of male hypogonadism. The IA used for E2 measurements showed poor correlation with MS and may only be suitable for the detection of high E2 in men.

Publisher

Bioscientifica

Subject

Endocrinology,General Medicine,Endocrinology, Diabetes and Metabolism

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3