Affiliation:
1. Tata Institute of Fundamental Research Hyderabad , 36/P Gopanpally, Hyderabad 500046, Telangana, India
Abstract
We describe the design, characterization, and application of a simple, highly collimated, and compact atomic/molecular beam source. This source is based on a segmented capillary design, constructed using a syringe needle. Angular width measurements and free molecular flow simulations show that the segmented structure effectively suppresses atoms traveling in off-axis directions, resulting in a narrow beam of Helium atoms having a width of 7 mrad (full width half maximum). We demonstrate an application of this source by using it for monitoring real-time changes in surface coverage on a clean Cu(110) surface exposed to oxygen by measuring the specular reflectivity of the Helium beam generated using this source.
Funder
Science and Engineering Research Board
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献