Affiliation:
1. Facility for Rare Isotope Beams, Michigan State University , East Lansing, Michigan 48824, USA
Abstract
At the ReAccelerator within the Facility for Rare Isotope Beams, a combination of an interchangeable aluminum foil and a silicon detector was developed to quantify isobaric contamination in rare isotope beams. The device is simple to operate and is now used routinely. In this article, we describe the system and show an application of the device to determine the level of contamination of an Si-32 rare isotope beam by stable S-32. In addition, we describe how the new diagnostic device helped confirm an enhancement of the beam purity prior to beam delivery to experiments.
Funder
U.S. Department of Energy
Reference22 articles.
1. Commissioning and first accelerated beams in the ReAccelerator (ReA3) of the National Superconducting Cyclotron Laboratory, MSU,2016
2. Gas stopping and reacceleration techniques at the Facility for Rare Isotope Beams (FRIB);Nucl. Instrum. Methods Phys. Res., Sect. B,2023
3. Advances of the FRIB project;Int. J. Mod. Phys. E,2019
4. Accelerator commissioning and rare isotope identification at the Facility for Rare Isotope Beams;Mod. Phys. Lett. A,2022
5. FRIB website, https://frib.msu.edu/; accessed 20 June 2023.