Affiliation:
1. School of Microelectronics, Southern University of Science and Technology 1 , Shenzhen 518055, China
2. Academy for Advanced Interdisciplinary Studies, Southern University of Science and Technology 2 , Shenzhen 518055, China
Abstract
The Ångström method is a promising thermal diffusivity measurement method for microfibers. Based on the heat-loss Ångström method and micro-Ångström method, the FFT (Fast Fourier Transform) Ångström method can be used to shorten the testing time and enhance the testing robustness of thermal conductivity. The FFT Ångström uses multi-frequency superimposed input signals and then uses the FFT for frequency division analysis. Since different frequencies are input at the same time, the measuring time is greatly reduced. Moreover, because different frequencies are input in the same environment, the random and environmental perturbations of the frequencies are the same, enhancing the fitting robustness. In contrast, the single feeding strategy is relatively time-consuming, and its measurement homogeneity for different frequencies cannot be guaranteed. By comparing the measurement results from a multi-frequency input and separated single-frequency input, the FFT shows good feasibility and robustness. It may also have great potential in other types of thermal wave measurements.
Funder
National Science Foundation for Excellent Young Scholars
National Natural Science Foundation of China
Technology Innovation Commission of Shenzhen
NSQKJJ