Analytical bias of automated immunoassays for six serum steroid hormones assessed by LC-MS/MS

Author:

Debeljak Željko1,Marković Ivana1,Šerić Vatroslav1,Horvat Vesna1,Mandić Sanja1,Mandić Dario1,Lukić Iva1,Pavela Jasna1

Affiliation:

1. Clinical Institute of Laboratory Diagnostics, Osijek University Hospital, Osijek, Croatia; Faculty of Medicine, JJ Strossmayer University of Osijek, Osijek, Croatia

Abstract

Introduction: There is a growing amount of evidence showing the significant analytical bias of steroid hormone immunoassays, but large number of available immunoassays makes conduction of a single comprehensive study of this issue hardly feasible. Aim of this study was to assess the analytical bias of six heterogeneous immunoassays for serum aldosterone, cortisol, dehydroepiandrosterone sulphate (DHEAS), testosterone, 17-hydroxyprogesterone (OHP) and progesterone using the liquid chromatography coupled to the tandem mass spectrometry (LC-MS/MS). Materials and methods: This method comparison study included 49 serum samples. Testosterone, DHEAS, progesterone and cortisol immunoassays were performed on the Abbott Architect i2000SR or Alinity i analysers (Abbott Diagnostics, Chicago, USA). DiaSorin’s Liaison (DiaSorin, Saluggia, Italy) and DIAsource’s ETI-Max 3000 analysers (DIAsource ImmunoAssays, Louvain-La-Neuve, Belgium) were chosen for aldosterone and OHP immunoassay testing, respectively. All immunoassays were evaluated against the LC-MS/MS assay relying on the commercial kit (Chromsystems, Gräfelfing, Germany) and LCMS-8050 analyser (Shimadzu, Kyoto, Japan). Analytical biases were calculated and method comparison was conducted using weighted Deming regression analysis. Results: Depending on the analyte and specific immunoassay, mean relative biases ranged from -31 to + 137%. Except for the cortisol, immunoassays were positively biased. For none of the selected steroids slope and intercept 95% confidence intervals simultaneously contained 0 and 1, respectively. Conclusions: Evaluated immunoassays failed to satisfy requirements for methods’ comparability and produced significant analytical biases in respect to the LC-MS/MS assay, especially at low concentrations.

Publisher

Croatian Society for Medical Biochemistry and Laboratory Medicine

Subject

Biochemistry, medical,Clinical Biochemistry

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3