Analytical evaluation of the new Seal Autoanalyzer 3 High Resolution for urinary iodine determination

Author:

Vidranski Valentina1,Franceschi Maja1,Krilić Dražena2,Jukić Tomislav1,Mihaljević Ivan3,Kusić Zvonko4

Affiliation:

1. Department for Oncology and Nuclear Medicine, University Hospital Center Sestre milosrdnice, Zagreb, Croatia; Faculty of medicine Osijek, Josip Juraj Strossmayer University of Osijek, Osijek, Croatia

2. Department for Oncology and Nuclear Medicine, University Hospital Center Sestre milosrdnice, Zagreb, Croatia

3. Department for Nuclear Medicine and Radiation Protection, University Hospital Center Osijek, Osijek, Croatia; Faculty of medicine Osijek, Josip Juraj Strossmayer University of Osijek, Osijek, Croatia

4. Croatian Academy of Sciences and Arts, Zagreb, Croatia

Abstract

Introduction: The aim of the study was to evaluate the analytical performance of the new colorimetric, automatic analyser, Seal AutoAnalyzer 3 High Resolution (Seal AA3 HR) (Seal Analytical, Wisconsin, USA) for urinary iodine measurement. Materials and methods: This study included testing of several analytical features of the method involving: imprecision (within-run %CVr, between- run %CVb and total laboratory precision %CVl), measurement uncertainty, carryover, linearity and method comparison, with 70 urine samples including the measuring range (20 - 700 μg/L). Results: Within-run, %CVb and %CVl of two control levels were 2.03% and 3.04%, 0.51% and 2.61%, and 2.09% and 4.01%, respectively. Carryover effect was less than 1%. The linearity was good in the range of urinary iodine values between 60 and 500 μg/L (R2 = 0.99). Good agreement of urinary iodine values was found between manual technique and Seal AA3 HR, using Passing-Bablok regression (y = 7.84 (- 3.00 to 15.29) + 0.95 (0.90 to 1.00) x) and Blant-Altman test. Cusum test for linearity indicates that there is no significant deviation from linearity (P > 0.1). Conclusions: The obtained results proved excellent precision, reproducibility and linearity, comparable to the already used, manual method. The New Seal AA3 HR automatic analyser is acceptable for urinary iodine measurement with very good analytical characteristics and can be used for urinary iodine epidemiological studies of the Croatian population.

Publisher

Croatian Society for Medical Biochemistry and Laboratory Medicine

Subject

Biochemistry, medical,Clinical Biochemistry

Reference29 articles.

1. World Health Organization (WHO). Iodine status worldwide: WHO global database on iodine deficiency. Geneva: WHO; 2004. Available at: http://whqlibdoc.who.int/publications/2004/9241592001.pdf. Accessed February 14th 2019.

2. Eastman CJ, Zimmermann MB. The Iodine Deficiency Disorders. [Updated 2018 Feb 6]. In: Feingold KR, Anawalt B, Boyce A, et al., eds. Endotext [Internet]. Available at: https://www.ncbi.nlm.nih.gov/books/NBK285556/. Accessed February 12th 2019.

3. Laurberg P, Pedersen IB, Carlé AS, editors. The U-Shaped Curve of Iodine Intake and Thyroid Disorders. In: Comprehensive Handbook of Iodine. Preedy VR, Burrow GN, Watson R, eds. Available at: https://www.sciencedirect.com/book/9780123741356/comprehensivehandbook-of-iodine. Accessed August 15th 2018. https://doi.org/10.1016/b978-0-12-374135-6.00047-9

4. The assessment of iodine status – populations, individuals and limitations.;Wainwright;Ann Clin Biochem,2019

5. Iodine excess as an environmental risk factor for autoimmune thyroid disease.;Luo;Int J Mol Sci,2014

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3