IntroductionIcterus, if not detected, can affect the validity of results delivered by clinical laboratories, leading to erroneous results. This study aims to define bilirubin interference for some biochemical analytes and compare it with the manufacturer’s data. Material and methodsSerum pools prepared with outpatients’ samples were spiked with increasing bilirubin concentration (Merck, reference14370, Darmstadt, Germany) up to 513 µmol/L in order to evaluate the bias for the following biochemical analytes: creatinine (CREA), creatine kinase (CK), cholesterol (CHOL), gamma-glutamyltransferase (GGT), high-density lipoprotein cholesterol (HDL), and total protein (TP). For each analyte, six pools of different concentrations were prepared. Measurements were made employing Cobas 8000 analyser c702-502, Roche Diagnostics (Mannheim, Germany). This study employed a study procedure defined by the Spanish Society of Laboratory Medicine. ResultsObtained bilirubin concentrations producing a negative interference were 103 µmol/L for CHOL, 205 µmol/L for TP and 410 µmol/L for CK, but only for CK values less than 100 U/L. Bilirubin concentrations lower than 513 µmol/L do not produce interference for HDL and GGT. Finally, for the studied bilirubin concentrations, there is no interference for CREA higher than 80 µmol/L. ConclusionIcterus interferences have been defined for each analyte, observing differences compared to data provided by the manufacturer. The evidence indicates that each laboratory should evaluate icteric interferences to ensure the high quality of the delivered results, thus benefiting patient care.

Publisher

Croatian Society for Medical Biochemistry and Laboratory Medicine

Subject

Biochemistry (medical),Clinical Biochemistry

Reference24 articles.

1. Evaluation of the interference due to hemoglobin, bilirubin and lipids on Immulite 2500 assays: a practical approach.;Steen;Ann Clin Biochem,2011

2. Interference testing.;Dimeski;Clin Biochem Rev,2008

3. Frequency of icteric interference in clinical chemistry laboratory tests and causes of severe icterus.;Mainali;Pract Lab Med,2021

4. Enzymatic determination of total serum cholesterol.;Allain;Clin Chem,1974

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3