IntroductionIcterus, if not detected, can affect the validity of results delivered by clinical laboratories, leading to erroneous results. This study aims to define bilirubin interference for some biochemical analytes and compare it with the manufacturer’s data.
Material and methodsSerum pools prepared with outpatients’ samples were spiked with increasing bilirubin concentration (Merck, reference14370, Darmstadt, Germany) up to 513 µmol/L in order to evaluate the bias for the following biochemical analytes: creatinine (CREA), creatine kinase (CK), cholesterol (CHOL), gamma-glutamyltransferase (GGT), high-density lipoprotein cholesterol (HDL), and total protein (TP). For each analyte, six pools of different concentrations were prepared. Measurements were made employing Cobas 8000 analyser c702-502, Roche Diagnostics (Mannheim, Germany). This study employed a study procedure defined by the Spanish Society of Laboratory Medicine.
ResultsObtained bilirubin concentrations producing a negative interference were 103 µmol/L for CHOL, 205 µmol/L for TP and 410 µmol/L for CK, but only for CK values less than 100 U/L. Bilirubin concentrations lower than 513 µmol/L do not produce interference for HDL and GGT. Finally, for the studied bilirubin concentrations, there is no interference for CREA higher than 80 µmol/L.
ConclusionIcterus interferences have been defined for each analyte, observing differences compared to data provided by the manufacturer. The evidence indicates that each laboratory should evaluate icteric interferences to ensure the high quality of the delivered results, thus benefiting patient care.
Publisher
Croatian Society for Medical Biochemistry and Laboratory Medicine
Subject
Biochemistry (medical),Clinical Biochemistry
Reference24 articles.
1. Evaluation of the interference due to hemoglobin, bilirubin and lipids on Immulite 2500 assays: a practical approach.;Steen;Ann Clin Biochem,2011
2. Interference testing.;Dimeski;Clin Biochem Rev,2008
3. Frequency of icteric interference in clinical chemistry laboratory tests and causes of severe icterus.;Mainali;Pract Lab Med,2021
4. Enzymatic determination of total serum cholesterol.;Allain;Clin Chem,1974
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献