Abstract
BackgroundCone-beam computed tomography (CBCT) imaging of the brain can be performed in the angiography suite to support various neurovascular procedures. Relying on CBCT brain imaging solely, however, still lacks full diagnostic confidence due to the inferior image quality compared with CT and various imaging artifacts that persist even with modern CBCT.ObjectiveTo perform a detailed evaluation of image artifact improvement using a new CBCT protocol which implements a novel dual-axis ‘butterfly’ trajectory.MethodsOur study included 94 scans from 47 patients who received CBCT imaging for assessment of either ischemia or hemorrhage during a neurovascular procedure. Both a traditional uni-axis ‘circular’ and novel dual-axis ‘butterfly’ protocol were performed on each patient (same-patient control). Each brain scan was divided into six regions and scored out of 3 based on six artifacts originating from various physics-based and patient-based sources.ResultsThe dual-axis trajectory produces CBCT images with significantly fewer image artifacts than the traditional circular scan (whole brain average artifact score, AS: 0.20 vs 0.33), with the greatest improvement in bone beam hardening (AS: 0.13 vs 0.78) and cone-beam artifacts (AS: 0.04 vs 0.55).ConclusionsRecent developments in CBCT imaging protocols have significantly improved image artifacts, which has improved diagnostic confidence for stroke and supports a direct-to-angiography suite transfer approach for patients with acute ischemic stroke.
Subject
Neurology (clinical),General Medicine,Surgery
Cited by
8 articles.