Abstract
BackgroundRaised intracranial pressure (ICP) contributes to approximately 20% of the admissions in the paediatric intensive care unit (PICU) in our setting. Timely identification and treatment of raised ICP is important to prevent brain herniation and death in such cases. The objective of this study was to examine the role of optic nerve sheath diameter (ONSD) in detecting clinically relevant raised ICP in children.MethodsA hospital-based observational analytical study in a PICU of a tertiary care institute in India on children aged 2–14 years. ONSD was measured in all children on three time points that is, day 1, day 2 and between day 4 and 7 of admission. ONSD values were compared between children with and without clinical signs of raised ICP.ResultsOut of 137 paediatric patients recruited, 34 had signs of raised ICP. Mean ONSD on day 1 was higher in children with signs of raised ICP (4.99±0.57 vs 4.06±0.40; p<0.01). Mean ONSD on day 2 also was higher in raised ICP patients (4.94±0.55 vs 4.04±0.40; p<0.01). The third reading between days 4 and 7 of admission was less than the first 2 values but still higher in raised ICP patients (4.48±1.26 vs 3.99±0.57; p<0.001). The cut-off ONSD value for detecting raised ICP was 4.46 mm on the ROC curve with an area under curve 0.906 (95% CI 0.844 to 0.968), 85.3% sensitivity and 86.4% specificity. There was no difference in ONSD between the right and the left eyes at any time point irrespective of signs of raised ICP.ConclusionWe found that measurement of ONSD by transorbital ultrasound was able to detect clinically relevant raised ICP with an excellent discriminatory performance at the cut-off value of 4.46 mm.