Author:
Roessingh A S de Buys,Drukker A,Guignard J-P
Abstract
AIMTo evaluate the reliability of dipstick measurements of urine specific gravity (U-SG).METHODSFresh urine specimens were tested for urine pH and osmolality (U-pH, U-Osm) by a pH meter and an osmometer, and for U-SG by three different methods (refractometry, automatic readout of a dipstick (Clinitek-50), and (visual) change of colour of the dipstick).RESULTSThe correlations between the visual U-SG dipstick measurements and U-SG determined by a refractometer and the comparison of Clinitek®-50 dipstick U-SG measurements with U-Osm were less than optimal, showing very wide scatter of values. Only the U-SG refractometer values and U-Osm had a good linear correlation. The tested dipstick was unreliable for the bedside determination of U-SG, even after correction for U-pH, as recommended by the manufacturer.CONCLUSIONSAmong the bedside determinations, only refractometry gives reliable U-SG results. Dipstick U-SG measurements should be abandoned.
Subject
Pediatrics, Perinatology and Child Health
Cited by
44 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献