Author:
Hasegawa Masahiro,Naito Yohei,Tone Shine,Sudo Akihiro
Abstract
Abstract
Background
Augmented reality (AR) provides the surgeon with direct visualization of radiological images by overlaying them on the patient. This study aimed to evaluate the accuracy of cup placement using a computed tomography (CT)-based AR navigation system.
Methods
Sixty-five prospectively enrolled patients underwent primary cementless total hip arthroplasty (THA) in a supine position using this novel AR navigation system, and changes in pelvic flexion angle (PFA) were evaluated. Absolute navigation errors were defined as the absolute differences between angles in the intraoperative navigation record and those measured on postoperative CT. Factors affecting the absolute navigation error in cup alignment were determined.
Results
Mean absolute change in PFA between preoperative CT and reduction was 2.1° ± 1.6°. Mean absolute navigation errors were 2.5° ± 1.7° in radiographic inclination (RI) and 2.5° ± 2.2° in radiographic anteversion (RA). While no factors significantly affecting absolute navigation error were found for RI, absolute change in PFA between preoperative CT and reduction correlated significantly with the absolute navigation error for RA.
Conclusion
This CT-based navigation system with AR enabled surgeons to place the cup more accurately than was possible by freehand placement during THA in a supine position.
Publisher
Springer Science and Business Media LLC
Subject
Orthopedics and Sports Medicine,Surgery
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献