Author:
Zhao Jingwei,Liu Yajun,Zhang Qi,Fan Mingxing,Han Xiaoguang,He Da,Liu Bo,Tian Wei
Abstract
Abstract
Background
This study aimed to evaluate the safety and efficacy of robot-assisted percutaneous pars–pedicle screw fixation surgery for treating Hangman’s fracture.
Methods
The study involved 33 patients with Hangman’s fracture who underwent robot-assisted fixation surgery using cannulated pars–pedicle screws through a percutaneous approach. The primary parameter evaluated was the accuracy of the screws according to the Gertzbein–Robbins scale, using postoperative CT images. Secondary parameters included the duration of surgery, intraoperative blood loss, postoperative hospital stay, and neurovascular injury.
Results
A total of 60 pars–pedicle screws were placed in 33 patients. Based on the Levine and Edwards classification, the patients included 12 cases of type I, 15 cases of type II, five cases of type IIa, and one atypical case. The average operative time was 92.4 ± 37.4 min, and the average blood loss was 22.4 ± 17.9 ml. Fifty-five of 60 screws were successfully placed within the bone. No screw-related neurovascular injury was observed, and satisfactory reduction was achieved in all cases.
Conclusion
Robot-assisted percutaneous pars–pedicle screw fixation is a safe and feasible method for treating Hangman’s fracture.
Trial registration: The study was retrospectively registered and approved by our center’s institutional review board.
Funder
Beijing Natural Science Foundation
CAMS Innovation Fund for Medical Sciences
Nova Program of Beijing Jishuitan Hospital
Publisher
Springer Science and Business Media LLC
Subject
Orthopedics and Sports Medicine,Surgery
Reference12 articles.
1. Levine AM, Edwards C. The management of traumatic spondylolisthesis of the axis. J Bone Joint Surg Am. 1985;67(2):217–26.
2. ElMiligui Y, Koptan W, Emran I. Transpedicular screw fixation for type II Hangman’s fracture: a motion preserving procedure. Eur Spine J. 2010;19:1299–305.
3. Sugimoto Y, Ito Y, Shimokawa T, et al. Percutaneous screw fixation for traumatic spondylolisthesis of the axis using iso-C 3D fluoroscopy-assisted navigation (case report). Minim Invasive Neurosurg. 2010;53:83–5.
4. Kovari VZ, Josvai A, Csokay A. Transpedicular direct osteosynthesis of hangman’s fracture from a mini-open exposure as a less invasive procedure: a technical note. Trauma Case Rep. 2017;12:66–71.
5. Han X, Tian W, Liu Y, et al. Safety and accuracy of robot-assisted versus fluoroscopy-assisted pedicle screw insertion in thoracolumbar spinal surgery: a prospective randomized controlled trial. J Neurosurg Spine. 2019;8:1–8.
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献