Abstract
Abstract
Background
Videolaryngoscopes with an operating channel may improve the intubation success rate in critically ill patients. We aimed to compare four channelled videolaryngoscopes to the Macintosh laryngoscope used for intubation of a high-fidelity simulation mannikin, in a scenario that simulated critical illness due to acute respiratory failure.
Results
Of the 79 residents who participated, 54 were considered inexperienced with orotracheal intubation. Each participant used all five devices in random order. The first-pass success rate was 97.5% [95% CI 91.1–99.7] for Airtraq™, KingVision™, and Pentax AWS200™, 92.4% [95% CI 84.2–97.2] for VividTrac VT-A100™, and 70.9% [95% CI 59.6–80.6] for direct Macintosh laryngoscopy. The first-pass success rate was significantly lower with direct Macintosh laryngoscopy than with the videolaryngoscopes (p < 0.0001 for Airtraq™, KingVision™, Pentax AWS200™, and VividTrac VT-A100™).
Conclusion
The Airtraq™, KingVision™, and Pentax AWS200™ channelled videolaryngoscopes produced high first-pass success rates with a lower boundary of the 95% CI above 90%. A multicentre, randomised controlled clinical study comparing channelled videolaryngoscopy to direct laryngoscopy should include one of these three videolaryngoscopes.
Publisher
Springer Science and Business Media LLC
Subject
Critical Care and Intensive Care Medicine
Cited by
7 articles.