Abstract
Abstract
Background
The aim of the present study was to develop and validate an inductively coupled plasma optical emission spectroscopic (ICP–OES) method for quantification of elemental impurities, i.e., Lead, Palladium, and Zinc, in voriconazole drug substance, and this method was employed for the regular sample analysis of Lead, Palladium, and Zinc in voriconazole drug substance for pharmaceutical use. The method has been validated using RF power of 1150 W, auxiliary gas of 0.5 L/min, and nebulizer flow of 0.4 L/min and plasma view at axial mode for Lead and Palladium and radial mode for Zinc. The wavelength was monitored for Lead, Palladium, and Zinc at 220.3 nm, 340.4 nm, and 213.8 nm respectively.
Results
The method is selective and is capable of detecting desired elemental impurities with regulatory acceptance limits in the presence of other elements. The validation experiments involve the demonstration of system suitability, specificity, LOD and LOQ, linearity, precision, and accuracy experiments. The linearity results obtained ˃ 0.9990 for all three impurities.
Conclusion
The proposed method is simple, sensitive quality control tool for the simultaneous quantitative determination of Lead, Palladium, and Zinc at low levels in voriconazole drug substance.
Graphical abstract
Publisher
Springer Science and Business Media LLC