Background
The aim of this study was to compare the performance of the new flow cytometer UF-5000 with the UF-1000i and Gram staining for determining bacterial patterns in urine samples.
Methods
Women who attended our clinic with symptoms suggestive of urinary tract infection were enrolled in the study. Mid-stream urine samples were collected for gram staining, urine analysis and urine cultures. Bacterial patterns were classified using the UF-1000i (none, cocci bacteria or rods/mixed growth), the UF-5000 (none, cocci, rods or mixed growth) and Gram staining.
Results
Among the 102 included samples, there were 10 g-positive cocci, 2 g-positive bacilli, 66 g-negative rods, and 24 mixed growth. The sensitivity/specificity of the UF-1000i was 81.8/91.1% for gram-negative rods and 23.5/96.9% for cocci/mixed. The sensitivity/specificity of the UF-5000 was 80.0/88.2% for gram negative rods and 70.0/86.5% for gram-positive cocci.
Conclusions
The UF-5000 demonstrated good sensitivity and specificity for Gram-negative bacilli and demonstrated an improved sensitivity for detecting Gram-positive cocci compared with the UF-1000i.