Abstract
Abstract
Objectives
To assess whether a deep learning-based system (DLS) with black-blood imaging for brain metastasis (BM) improves the diagnostic workflow in a multi-center setting.
Materials and methods
In this retrospective study, a DLS was developed in 101 patients and validated on 264 consecutive patients (with lung cancer) having newly developed BM from two tertiary university hospitals, which performed black-blood imaging between January 2020 and April 2021. Four neuroradiologists independently evaluated BM either with segmented masks and BM counts provided (with DLS) or not provided (without DLS) on a clinical trial imaging management system (CTIMS). To assess reading reproducibility, BM count agreement between the readers and the reference standard were calculated using limits of agreement (LoA). Readers’ workload was assessed with reading time, which was automatically measured on CTIMS, and were compared between with and without DLS using linear mixed models considering the imaging center.
Results
In the validation cohort, the detection sensitivity and positive predictive value of the DLS were 90.2% (95% confidence interval [CI]: 88.1–92.2) and 88.2% (95% CI: 85.7–90.4), respectively. The difference between the readers and the reference counts was larger without DLS (LoA: −0.281, 95% CI: −2.888, 2.325) than with DLS (LoA: −0.163, 95% CI: −2.692, 2.367). The reading time was reduced from mean 66.9 s (interquartile range: 43.2–90.6) to 57.3 s (interquartile range: 33.6–81.0) (P <.001) in the with DLS group, regardless of the imaging center.
Conclusion
Deep learning-based BM detection and counting with black-blood imaging improved reproducibility and reduced reading time, on multi-center validation.
Funder
Ministry of Health and Welfare
RINK-CR
National Research Foundation of Korea
Publisher
Springer Science and Business Media LLC
Cited by
1 articles.
订阅此论文施引文献