Functionalization of screw implants with superelastic structured Nitinol anchoring elements

Author:

Hamann IsabellORCID,Schleifenbaum Stefan,Rotsch Christian,Drossel Welf-Guntram,Heyde Christoph-Eckhard,Leimert Mario

Abstract

Abstract Background Demographic change is leading to an increase in the number of osteoporotic patients, so a rethink is required in implantology in order to be able to guarantee adequate anchoring stability in the bone. The functional modification of conventional standard screw implants using superelastic, structured Ti6Al4V anchoring elements promises great potential for increasing anchoring stability. Methods For this purpose, conventional screw implants were mechanically machined and extended so that structured-superelastic-positionable-Ti6Al4V anchoring elements could be used. The novel implants were investigated with three tests. The setup of the anchoring elements was investigated in CT studies in an artificial bone. In a subsequent simplified handling test, the handling of the functional samples was evaluated under surgical conditions. The anchorage stability compared to standard screw implants was investigated in a final pullout test according to ASTM F543—the international for the standard specification and test methods for metallic medical bone screws. Results The functionalization of conventional screw implants with structured superelastic Ti6Al4V anchoring elements is technically realizable. It was demonstrated that the anchoring elements can be set up in the artificial bone without any problems. The anchorage mechanism is easy to handle under operating conditions. The first simplified handling test showed that at the current point of the investigations, the anchoring elements have no negative influence on the surgical procedure (especially under the focus of screw implantation). Compared to conventional standard screws, more mechanical work is required to remove the functional patterns completely from the bone. Conclusion In summary, it was shown that conventional standard screw implants can be functionalized with Ti6Al4V-structured NiTi anchoring elements and the new type of screws are suitable for orthopedic and neurosurgical use. A first biomechanical test showed that the anchoring stability could be increased by the anchoring elements.

Funder

bundesministerium für bildung und forschung

Fraunhofer-Institut für Werkzeugmaschinen und Umformtechnik IWU

Publisher

Springer Science and Business Media LLC

Subject

Radiology, Nuclear Medicine and imaging,Biomedical Engineering,General Medicine,Biomaterials,Radiological and Ultrasound Technology

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3