Author:
Zhang Qiuwang,Duncan Sebastian,Szulc Daniel A.,de Mestral Charles,Kutryk Michael JB
Abstract
Abstract
Background
Thrombosis is a common cause of vascular prosthesis failure. Antibody coating of prostheses to capture circulating endothelial progenitor cells to aid endothelialization on the device surface appears a promising solution to prevent thrombus formation. Compared with random antibody immobilization, oriented antibody coating (OAC) increases antibody-antigen binding capacity and reduces antibody immunogenicity in vivo. Currently, few OAC methods have been documented, with none possessing clinical application potential.
Results
Dopamine and the linker amino-PEG8-hydrazide-t-boc were successfully deposited on the surface of cobalt chromium (CC) discs, CC stents and expanded polytetrafluoroethylene (ePTFE) grafts under a slightly basic condition. CD34 antibodies were immobilized through the reaction between aldehydes in the Fc region created by oxidation and hydrazides in the linker after t-boc removal. CD34 antibody-coated surfaces were integral and smooth as shown by scanning electron microscopy (SEM), had significantly reduced or no substrate-specific signals as revealed by X-ray photoelectron spectroscopy, were hospitable for HUVEC growth as demonstrated by cell proliferation assay, and specifically bound CD34 + cells as shown by cell binding testing. CD34 antibody coating turned hydrophobic property of ePTFE grafts to hydrophilic. In a porcine carotid artery interposition model, a confluent monolayer of cobblestone-shaped CD31 + endothelial cells on the luminal surface of the CD34 antibody coated ePTFE graft were observed. In contrast, thrombi and fibrin fibers on the bare graft, and sporadic cells on the graft coated by chemicals without antibodies were seen.
Conclusion
A universal, OAC method was developed. Our in vitro and in vivo data suggest that the method can be potentially translated into clinical application, e.g., modifying ePTFE grafts to mitigate their thrombotic propensity and possibly provide for improved long-term patency for small-diameter grafts.
Funder
the Alayne Metrick Fund, Toronto, ON, Canada.
Publisher
Springer Science and Business Media LLC
Subject
Cell Biology,Molecular Biology,Biomedical Engineering,Environmental Engineering
Reference48 articles.
1. Prager M, Polterauer P, Böhmig HJ, et al. Collagen versus gelatin-coated Dacron versus stretch polytetrafluoroethylene in abdominal aortic bifurcation graft surgery: results of a seven-year prospective, randomized multicenter trial. Surgery. 2001;130:408–14.
2. Vartanian SM, Conte MS. Surgical intervention for peripheral arterial disease. Circ Res. 2015;116:1614–28.
3. Halbert RJ, Nicholson G, Nordyke RJ, Pilgrim A, Niklason L. Patency of ePTFE arteriovenous graft placements in hemodialysis patients: systematic literature review and meta-analysis. Kidney 360. 2020;1:1437–46.
4. Hehrlein FW, Schlepper M, Loskot F, Scheld HH, Walter P, Mulch J. The use of expanded polytetrafluoroethylene (PTFE) grafts for myocardial revascularization. J Cardiovasc Surg (Torino). 1984;25:549–53.
5. Chard RB, Johnson DC, Nunn GR, Cartmill TB. Aorta-coronary bypass grafting with polytetrafluoroethylene conduits. Early and late outcome in eight patients. J Thorac Cardiovasc Surg. 1987;94:132–4.