Abstract
Abstract
Background
To date, measurement of intracellular cystine is used for the therapeutic monitoring of patients affected by cystinosis in treatment with cysteamine. Since this method is time and sample consuming, development of a faster method to quantify cysteamine would be extremely useful in order to help clinicians to adjust dosages of cysteamine and to define better the pharmacokinetic profile of this drug. The aim of the study was to develop a liquid chromatography tandem mass spectrometry method for the quantification of cysteamine in plasma samples and to test its applicability on plasma samples derived from patients with nephropathic infantile cystinosis in treatment with cysteamine.
Results
The percentage of accuracy of the developed method varied between 97.80 and 106.00% and CV% between 0.90 and 6.93%. There was no carry over. The calibration curves were built from 2.5 to 50 µM. The limit of detection and the lower limit of quantification occurred at 0.25 and 1.25 µM respectively. Cysteamine was stable up to 2 months at -20 °C. Concentrations of cysteamine and intracellular cystine of 4 patients were in line with data previously reported.
Conclusion
The proposed method showed an appropriate selectivity, specificity, linearity, sensibility, accuracy, precision and good applicability to samples.
Publisher
Springer Science and Business Media LLC
Subject
Pharmacology (medical),Genetics (clinical),General Medicine
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献