Abstract
Abstract
Background
The 4th National Audit Project of The Royal College of Anaesthetists and The Difficult Airway Society (NAP4) reported a higher incidence of supraglottic airway device (SAD) related pulmonary aspiration in obese patients especially with the first-generation SADs. The latest single-use SAD, the Protector™ provides a functional separation of the respiratory and digestive tracts and its laryngeal cuff with two ports allowing additional suction in tandem with the insertion of a gastric tube. The laryngeal cuff of LMA Protector™ allows a large catchment reservoir in the event of gastric content aspiration.
Methods
We evaluated the performance characteristics of the LMA Protector™ in 30 unparalysed, moderately obese patients. First attempt insertion rate, time for insertion, oropharyngeal leak pressure (OLP), and incidence of complications were recorded.
Results
We found high first and second attempt insertion rates of 28(93%) and 1(33%) respectively, with one failed attempt where no capnography trace could be detected, presumably from a downfolded device tip. The LMA Protector™ was inserted rapidly in 21.0(4.0) seconds and demonstrated high OLP of 31.8(5.4) cmH2O. Fibreoptic assessment showed a clear view of vocal cords in 93%. The incidence of blood staining on removal of device was 48%, postoperative sore throat 27%, dysphagia 10% and dysphonia 20% (all self-limiting, resolving a few hours postoperatively).
Conclusions
We conclude that the LMA Protector™ was associated with easy, expedient first attempt insertion success, demonstrating high oropharyngeal pressures and good anatomical position in the moderately obese population, with relatively low postoperative airway morbidity.
Trial registration
Australian New Zealand Clinical Trials Registry, ACTRN12617001152314. Registered 7 August 2017.
Publisher
Springer Science and Business Media LLC
Subject
Anesthesiology and Pain Medicine
Reference19 articles.
1. Cook TM, Woodall N, Frerk C. Major complications of airway management in the UK: results of the fourth National Audit Project of the Royal College of Anaesthetists and the difficult airway society. Part 1: anaesthesia. Br J Anaesth. 2011;106(5):617–31.
2. LMA Protector Instruction for use, Teleflex, 2015 [http://www.lmacoifu.com/sites/default/files/node/1928/ifu/revision/3285/pbe2100000b-lma-protector-ifuuk.pdf].
3. Brimacombe J, Berry AJA. A proposed fiber-optic scoring system to standardize the assessment of laryngeal mask airway position. Analgesia. 1993;76(2):457.
4. Sorbello M, Petrini FJTJA. Supraglottic airway devices: the search for the best insertion technique or the time to change our point of view? Reanimation. 2017;45(2):76.
5. Daniel WW, Cross CL. Biostatistics: a foundation for analysis in the health sciences. New York: Wiley; 2018.
Cited by
6 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献