Abstract
Abstract
Background
To introduce a new method for measurement of surface roughness of the endodontic instrument, before and after instrumentation, using the Field Emission Scanning Electronic Microscope (FE-SEM) combined with the ImageJ software.
Methods
Twenty J-shape resin blocks were divided into two groups, ten blocks of each group. Simulated root canal inside the resin block was 16 mm length, 600 angle of curvature, and radius of 4.5 mm. Ten WaveOne Gold Primary and 10 Reciproc Blue R25 instruments were used for root canal instrumentation. The instruments were scanned before and after instrumentation with special molds made to ensure the same areas at the point located 3 mm from the tips of the instruments using the FE-SEM. These scanned images were analyzed using the ImageJ. The arithmetical mean roughness (Ra), root mean square roughness (Rq), and the average distance between the highest peak and lowest valley in each sampling length (Rz) were calculated by ImageJ for quantitative analyses. The paired-t test was performed to analyze the data using the SPSS 22.0 at the significance of .05.
Results
Almost all surface roughness values were decreased. However, these decreases were not statistically significant (P > .05).
Conclusions
The FE-SEM combined with the ImageJ was the reliable and appropriate modality for measurement surface roughness of instruments.
Publisher
Springer Science and Business Media LLC
Reference28 articles.
1. Pham K, Phan T. Evaluation of root canal preparation using two nickel-titanium instrument systems via cone-beam computed tomography. Saudi Endodontic J. 2019;9(3):210–5.
2. Pham K, Nguyen N. Cutting efficiency and dentinal defects using two single-file continuous rotary nickel-titanium instruments. Saudi Endodontic J. 2020;10(1):56–60.
3. Ha J-H, Sigurdsson A, De-Deus G, Versluis A, Kwak SW, Kim H-C. Torsional behavior of WaveOne gold endodontic file with the dedicated motor of the original WaveOne file. Materials (Basel). 2018;11(7):1150.
4. De-Deus G, Cardoso ML, Belladonna FG, Cavalcante DM, Simões-Carvalho M, Souza EM, Lopes RT, Silva EJNL. Performance of reciproc blue R25 instruments in shaping the canal space without glide path. J Endod. 2019;45(2):194–8.
5. Gambarini G, Galli M, Di Nardo D, Seracchiani M, Donfrancesco O, Testarelli L. Differences in cyclic fatigue lifespan between two different heat treated NiTi endodontic rotary instruments: WaveOne gold vs EdgeOne fire. J Clin Exp Dent. 2019;11(7):e609–13.
Cited by
18 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献