Affiliation:
1. Auckland Bioengineering Institute and
2. Departments of 2Engineering Science and
3. Physiology, The University of Auckland, Auckland, New Zealand
Abstract
We describe a unique work-loop calorimeter with which we can measure, simultaneously, the rate of heat production and force-length work output of isolated cardiac trabeculae. The mechanics of the force-length work-loop contraction mimic those of the pressure-volume work-loops experienced by the heart. Within the measurement chamber of a flow-through microcalorimeter, a trabecula is electrically stimulated to respond, under software control, in one of three modes: fixed-end, isometric, or isotonic. In each mode, software controls the position of a linear motor, with feedback from muscle force, to adjust muscle length in the desired temporal sequence. In the case of a work-loop contraction, the software achieves seamless transitions between phases of length control (isometric contraction, isometric relaxation, and restoration of resting muscle length) and force control (isotonic shortening). The area enclosed by the resulting force-length loop represents the work done by the trabecula. The change of enthalpy expended by the muscle is given by the sum of the work term and the associated amount of evolved heat. With these simultaneous measurements, we provide the first estimation of suprabasal, net mechanical efficiency (ratio of work to change of enthalpy) of mammalian cardiac trabeculae. The maximum efficiency is at the vicinity of 12%.
Publisher
American Physiological Society
Subject
Physiology (medical),Physiology
Cited by
50 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献