Abstract
System matrix computation for iterative reconstruction algorithms in SPECT based on direct measurements
A method for system matrix calculation in the case of iterative reconstruction algorithms in SPECT was implemented and tested. Due to a complex mathematical description of the geometry of the detector set-up, we developed a method for system matrix computation that is based on direct measurements of the detector response. In this approach, the influence of the acquisition equipment on the image formation is measured directly. The objective was to obtain the best quality of reconstructed images with respect to specified measures. This is indispensable in order to be able to perform reliable quantitative analysis of SPECT images. It is also especially important in non-hybrid gamma cameras, where not all physical processes that disturb image acquisition can be easily corrected. Two experiments with an 131I point source placed at different distances from the detector plane were performed allowing the detector response to be acquired as a function of the point source distance. An analytical Gaussian function was fitted to the acquired data in both the one- and the two-dimensional case. A cylindrical phantom filled with a water solution of 131I containing a region of "cold" spheres as well as a uniform solution (without any spheres) was used to perform algorithm evaluation. The reconstructed images obtained by using four different of methods system matrix computation were compared with those achieved using reconstruction software implemented in the gamma camera. The contrast of the spheres and uniformity were compared for each reconstruction result and also with the ranges of those values formulated by the AAPM (American Association of Physicists in Medicine). The results show that the implementation of the OSEM (Ordered Subsets Expectation Maximization) algorithm with a one-dimensional fit to the Gaussian CDR (Collimator-Detector Response) function provides the best results in terms of adopted measures. However, the fit of the two-dimensional function also gives satisfactory results. Furthermore, the CDR function has the potential to be applied to a fully 3D OSEM implementation. The lack of the CDR in system matrix calculation results in a very noisy image that cannot be used for diagnostic purposes.
Subject
Applied Mathematics,Engineering (miscellaneous),Computer Science (miscellaneous)
Reference19 articles.
1. Monte Carlo modeling of gamma cameras for I-131 imaging in targeted radiotherapy;D. Autret;Cancer Biotherapy and Radiopharmaceuticals,2005
2. Evaluation of SPECT-CT image fusion quality control;D. Borys;Biocybernetics and Biomedical Engineering,2006
3. Analytic and iterative reconstruction algorithms in SPECT;P. Bruyant;Journal of Nuclear Medicine,2002
4. A method for attenuation correction in radionuclide computed tomography;L. Chang;IEEE Transactions on Nuclear Science,1978
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献