Is Valgus Cut Angle Based on Radiographic Measurements in Total Knee Arthroplasty Really Inaccurate? A Comparison of Two- and Three-Dimensional Measurements

Author:

Liu LiMing1ORCID,Lei Kai1ORCID,Chen Xin1,Fan HuaQuan1,Yang Liu1,Guo Lin1ORCID

Affiliation:

1. Center for Joint Surgery, Southwest Hospital, Third Military Medical University, Chongqing, China

Abstract

AbstractRadiographs are widely used to measure distal femoral valgus cut angle (VCA) in total knee arthroplasty (TKA), but its accuracy is controversial. This study used three-dimensional (3D) reconstruction models to verify the accuracy of VCA measurements on radiographs, and explore the correlation of VCA with hip–knee–ankle (HKA) angle and lateral femoral bowing angle (FBA). A total of 444 osteoarthritis knees of 444 patients from August 2016 to June 2018 was included retrospectively. On radiographs, two-dimensional VCA (VCA-2D) was measured between the femoral mechanical axis and the distal femoral anatomical axis, and HKA was measured between the femoral mechanical axis and the tibial mechanical axis. On the coronal projection of computed tomography 3D models, the anatomical landmarks used for VCA-3D measurements were the same as those on the radiographs, FBA was measured between the proximal and distal femoral anatomical axis. The distributions of VCA-2D and VCA-3D were evaluated by means and variances. The correlation between HKA and VCA and between FBA and VCA was explored. There was a statistical difference between VCA-2D and VCA-3D (p < 0.001), but the deviation was very small (0.15 ± 0.69 degrees), 83.3% of the deviations were less than 1 degree. VCA would increase both in 2D and 3D with increasing of FBA and HKA varus. There was no statistically significant difference between VCA-2D and VCA-3D in patients with moderate varus knees (0–8 degrees of varus) and mild bowing femurs (FBA <5 degrees). Overall, the deviation caused by using radiography to measure VCA was negligible. VCA measurements using radiographs were accurate in patients with moderate varus knees and mildly bowed femurs. This study reflects level of evidence III.

Funder

National Key R&D Program of China

Guangdong Provincial Key R&D Program

Publisher

Georg Thieme Verlag KG

Subject

Orthopedics and Sports Medicine,Surgery

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3