Isolated Tibial Component Failure in Total Knee Arthroplasty: A Case Series Evaluating Inflammatory Response versus Mechanical Failure

Author:

Yokhana Sanar S.12,Bergum Christopher2,Ren Weiping3,Markel David C.4

Affiliation:

1. Department of Orthopaedic Surgery, Detroit Medical Center, Detroit, Michigan

2. Department of Orthopaedic Surgery, Providence-Providence Park Hospital, Southfield, Michigan

3. Department of Biomedical Engineering, Wayne State University, Detroit, Michigan

4. Department of Orthopaedic Surgery, St. John-Providence Hospital, Southfield, Michigan

Abstract

AbstractTotal knee prostheses are routinely redesigned to improve performance, longevity, and to closer mimic the native kinematics of the knee. Despite continued improvements, all knee implants, even those with proven design features, have failures. We identified a cohort of patients with isolated tibial component failures that occurred in a popular and successful knee system. Our purpose was to (1) characterize the observed radiographic failure pattern; (2) investigate the biologic response that may have contributed to the failure; and (3) to determine if the failure mechanism was of a biological or a mechanical nature. Twenty-one knees from 19 patients met the inclusion criteria of isolated tibial component failure in a single knee implant system. Radiographs from the primary and revision knee surgery were analyzed for implant positioning and the failure pattern. Inflammatory biomarkers interleukin (IL)-1β, IL-6, and tumor necrosis factor (TNF)-α were available in 16/21 knees and peripheral CD14+/16+ monocytes were measured in 14 of the aforementioned 16 knee revisions. Serum CD3, CD4, and CD19 were measured in 10 of the aforementioned 14 knees. Additionally, white blood cell (WBC) count, erythrocyte sedimentation rate (ESR), and C-reactive protein (CRP) were measured to rule out infection as a cause of the cytokine upregulation. Radiographic findings demonstrated that all of the 21 tibial components were implanted in either neutral or 2 to 3 degrees varus in the coronal plane, and none of the revisions was implanted in valgus. All tibias showed obvious radiographic loosening and the implant failed into varus. The inflammatory biomarkers IL-1β, IL-6, and TNF-α were negative. WBC, ESR, and CRP were normal. Serum CD3, CD4, and CD19 flow cytometry analyses were found to be in the normal range. Peripheral CD14+/16+ and total CD16+ monocytes measurements were consistent with previous findings of patients with osteoarthritis, rather than particulate-induced inflammatory loosening. The findings support the implant failure observed in our study occurred by a different mechanism than the wear debris–induced aseptic loosening. We believe that a mechanical failure can occur based on our findings. The loosening, collapse, and debonding from the cement may have been related to the implantation technique, stresses due to favorable rotational freedoms of the implant, or patient characteristics/behavior. Continued exploration into implant failure mechanisms and particularly into the biologic response associated with failure is ongoing.

Publisher

Georg Thieme Verlag KG

Subject

Orthopedics and Sports Medicine,Surgery

Cited by 4 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3