Two Recombinant Tissue Factor Reagents Compared to Conventional Thromboplastins for Determination of International Normalised Ratio: A Thirty-three-laboratory Collaborative Study

Author:

Kitchen S1,Jennings I1,Woods T A L1,Walker I D2,Preston F E1,

Affiliation:

1. The University Department of Haematology, Royal Hallamshire Hospital, Sheffield

2. The Department of Haematology, Glasgow Royal Infirmary, Glasgow, UK

Abstract

SummaryRecent advances in recombinant technology have led to the development of prothrombin time (PT) reagents containing recombinant tissue factor which has been lipidated to allow expression of procoagulant activity. In this study we have compared International Normalised Ratios (INRs) determined using two such reagents and conventional thromboplastins in widespread use in the UK.Lyophilised plasma samples from eight different warfarinised patients were distributed to 33 laboratories in the UK. Each participant determined prothrombin times on 20 local fresh normal plasmas (used to derive mean normal PT and calculate INR) and the eight lyophilised samples, using manual technique and the following thromboplastins; Recombiplastin (Ortho Diagnostics Ltd); Innovin (Baxter Diagnostics Ltd); the conventional thromboplastin in local use.For eight plasmas the mean INRs determined with different reagents were as follows: Innovin (33 laboratories) - 3.4; Manchester Reagent (MR = 8 laboratories) - 3.4; Recombiplastin (33 laboratories) - 3.7; Instrumentation Laboratory (IL = 13 laboratories) - 4.4.Mean INR results with Recombiplastin were on average 7% greater than those obtained with Innovin, 8% greater than results with MR and 18% less than INRs with IL thromboplastin. There was no significant difference between results obtained with Innovin and MR. In contrast INRs obtained with IL were markedly (mean 28%) greater than results obtained with Innovin.This study employed lyophilised plasma and it is possible that some of the relationships described are influenced by this. However, the lyo-philisation process employed did not influence the relationship between INRs of warfarinised plasmas obtained by the four main reagents described, indicating that the results are relevant to routine clinical practice.In conclusion, our data show some important differences are present between INRs determined using Recombiplastin, Innovin and two conventional thromboplastins.

Publisher

Georg Thieme Verlag KG

Subject

Hematology

Cited by 8 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3