Multi-Center Study of Thromboplastin Calibration Precision – Influence of Reagent Species, Composition, and International Sensitivity Index (ISI)

Author:

Besselaar A M H P van den1,Bertina R M1

Affiliation:

1. The Thrombosis and Haemostasis Research Unit, Department of Haematology, Leiden University Hospital, Leiden, The Netherlands

Abstract

SummaryFour thromboplastin reagents were tested by 18 laboratories in Europe, North-America, and Australasia, according to a detailed protocol. One thromboplastin was the International Reference Preparation for ox brain thromboplastin combined with adsorbed bovine plasma (coded OBT/79), and the second was a certified reference material for rabbit brain thromboplastin, plain (coded CRM 149R). The other two thromboplastin reagents were another rabbit plain brain thromboplastin (RP) with a lower ISI than CRM 149R and a rabbit brain thromboplastin combined with adsorbed bovine plasma (RC). Calibration of the latter two reagents was performed according to methods recommended by the World Health Organization (W. H. O.).The purpose of this study was to answer the following questions: 1) Is the calibration of the RC reagent more precise against the bovine/combined (OBT/79) than against the rabbit/plain reagent (CRM 149R)? 2) Is the precision of calibration influenced by the magnitude of the International Sensitivity Index (ISI)?The lowest inter-laboratory variation of ISI was observed in the calibration of the rabbit/plain reagent (RP) against the other rabbit/plain reagent (CRM 149R) (CV 1.6%). The highest interlaboratory variation was obtained in the calibration of rabbit/plain (RP) against bovine/combined (OBT/79) (CV 5.1%). In the calibration of the rabbit/combined (RC) reagent, there was no difference in precision between OBT/79 (CV 4.3%) and CRM 149R (CV 4.2%). Furthermore, there was no significant difference in the precision of the ISI of RC obtained with CRM 149R (ISI = 1.343) and the rabbit/plain (RP) reagent with ISI = 1.14. In conclusion, the calibration of RC could be performed with similar precision with either OBT/79 or CRM 149R, or RP.The mean ISI values calculated with OBT/79 and CRM 149R were practically identical, indicating that there is no bias in the ISI of these reference preparations and that these reference preparations have been stable since their original calibration studies in 1979 and 1987, respectively.International Normalized Ratio (INR) equivalents were calculated for a lyophilized control plasma derived from patients treated with oral anticoagulants. There were small but significant differences in the mean INR equivalents between the bovine and rabbit thromboplastins. There were no differences in the interlaboratory variation of the INR equivalents, when the four thromboplastins were compared.

Publisher

Georg Thieme Verlag KG

Subject

Hematology

Cited by 12 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3