A feasibility study to predict 3D dose delivery accuracy for IMRT using DenseNet with log files

Author:

Huang Ying123,Cai Ruxin3,Pi Yifei4,Ma Kui5,Kong Qing1,Zhuo Weihai2,Kong Yan6

Affiliation:

1. Institute of Modern Physics, Fudan University, Shanghai, China

2. Institute of Radiation Medicine, Fudan University, Shanghai, China

3. Shanghai Chest Hospital, School of Medicine, Shanghai Jiao Tong University, Shanghai, China

4. Department of Radiation Oncology, The First Affiliated Hospital of Zhengzhou University, Henan, China

5. Varian Medical Systems, Beijing, China

6. Department of Radiation Oncology, Affiliated Hospital of Jiangnan University, Jiangsu, China

Abstract

OBJECTIVE: This study aims to explore the feasibility of DenseNet in the establishment of a three-dimensional (3D) gamma prediction model of IMRT based on the actual parameters recorded in the log files during delivery. METHODS: A total of 55 IMRT plans (including 367 fields) were randomly selected. The gamma analysis was performed using gamma criteria of 3% /3 mm (Dose Difference/Distance to Agreement), 3% /2 mm, 2% /3 mm, and 2% /2 mm with a 10% dose threshold. In addition, the log files that recorded the gantry angle, monitor units (MU), multi-leaf collimator (MLC), and jaws position during delivery were collected. These log files were then converted to MU-weighted fluence maps as the input of DenseNet, gamma passing rates (GPRs) under four different gamma criteria as the output, and mean square errors (MSEs) as the loss function of this model. RESULTS: Under different gamma criteria, the accuracy of a 3D GPR prediction model decreased with the implementation of stricter gamma criteria. In the test set, the mean absolute error (MAE) of the prediction model under the gamma criteria of 3% /3 mm, 2% /3 mm, 3% /2 mm, and 2% /2 mm was 1.41, 1.44, 3.29, and 3.54, respectively; the root mean square error (RMSE) was 1.91, 1.85, 4.27, and 4.40, respectively; the Sr was 0.487, 0.554, 0.573, and 0.506, respectively. There was a correlation between predicted and measured GPRs (P < 0.01). Additionally, there was no significant difference in the accuracy between the validation set and the test set. The accuracy in the high GPR group was high, and the MAE in the high GPR group was smaller than that in the low GPR group under four different gamma criteria. CONCLUSIONS: In this study, a 3D GPR prediction model of patient-specific QA using DenseNet was established based on log files. As an auxiliary tool for 3D dose verification in IMRT, this model is expected to improve the accuracy and efficiency of dose validation.

Publisher

IOS Press

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3