Author:
Wieland D. C. F.,Zander T.,Garamus V. M.,Krywka C.,Dedinaite A.,Claesson P.,Willumeit-Römer R.
Abstract
A newly developed high-pressure rheometer forin situX-ray scattering experiments is described. A commercial rheometer was modified in such a way that X-ray scattering experiments can be performed under different pressures and shear. First experiments were carried out on hyaluronan, a ubiquitous biopolymer that is important for different functions in the body such as articular joint lubrication. The data hint at a decreased electrostatic interaction at higher pressure, presumably due to the increase of the dielectric constant of water by 3% and the decrease of the free volume at 300 bar.
Publisher
International Union of Crystallography (IUCr)
Subject
Instrumentation,Nuclear and High Energy Physics,Radiation
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献