Author:
Espeso José I.,Cloetens Peter,Baruchel José,Härtwig Jürgen,Mairs Trevor,Biasci Jean Claude,Marot Gérard,Salomé-Pateyron Murielle,Schlenker Michel
Abstract
The lateral coherence length is of the order of 100 µm at the `long' (145 m) ID19 beamline of the ESRF, which is mainly devoted to imaging. Most of the optical elements located along the X-ray path can thus act as `phase objects', and lead to spurious contrast and/or to coherence degradation, which shows up as an enhanced effective angular size of the source. Both the spurious contrast and the coherence degradation are detrimental for the images (diffraction topographs, tomographs, phase-contrast images) produced at this beamline. The problems identified and the way they were solved during the commissioning of ID19 are reported. More particularly, the role of the protection foils located in the front end, the beryllium windows, the filters and the monochromator defects (scratches, dust, small vibrations) is discussed.
Publisher
International Union of Crystallography (IUCr)
Subject
Instrumentation,Nuclear and High Energy Physics,Radiation
Cited by
32 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献