Author:
Rantataro Samuel,Ferrer Pascual Laura,Laurila Tomi
Abstract
AbstractIt is widely stated that ascorbic acid (AA) interferes with the electrochemical detection of neurotransmitters, especially dopamine, because of their overlapping oxidation potentials on typical electrode materials. As the concentration of AA is several orders of magnitude higher than the concentration of neurotransmitters, detection of neurotransmitters is difficult in the presence of AA and requires either highly stable AA concentration or highly selective neurotransmitter sensors. In contrast to the common opinion, we show that AA does not always interfere electrochemical detection of neurotransmitters. The decay of AA is rapid in cell culture medium, having a half-time of 2.1 hours, according to which the concentration decreases by 93% in 8 hours and by 99.75% in 18 hours. Thus, AA is eventually no longer detected by electrodes and the concentration of neurotransmitters can be effectively monitored. To validate this claim, we used unmodified single-wall carbon nanotube electrode to measure dopamine at physiologically relevant concentration range (25–1000 nM) from human midbrain organoid medium with highly linear response. Finally, AA is known to affect dopamine oxidation current through regeneration of dopamine, which complicates precise detection of small amounts of dopamine. By designing experiments as described here, this complication can be completely eliminated.
Funder
Horizon 2020 Framework Programme
Publisher
Springer Science and Business Media LLC
Cited by
10 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献