A promising approach in laser vitrectomy executed by plasma-mediated removal of vitreous body via a diode-pumped Q-switched Nd:YAG laser

Author:

Uthoff Daniel,Oravecz Rita,Kuehnl Romina,Rubin-Schwarz Friederike,Frentzen Matthias,Schrage Norbert,Meister Jörg

Abstract

AbstractTo examine the applicability of plasma-mediated vitreous body removal, a diode-pumped Q-switched Neodymium:YAG laser was used for a possible application in eye surgery/vitrectomy. On a total of 1500 porcine vitreous bodies, removal rates were evaluated by comparing different LaserVit-tip designs (Mark I/II Gauge 19 and Mark III Gauge 22). The Nd:YAG laser, operating at a wavelength of 1064 nm and a pulse duration of 4 ns, was utilized for vitreous body removal with respective settings of 2, 3 and 4 mJ and pulse repetition rates (cut rates) from 5 to 25 Hz (300–1500 /min) in 5 Hz-steps as well as for 100 Hz (6000 cuts/min). The exposure times were selected at 10, 20, 40 and 60 s, respectively. Comparative measurements were carried out with mechanical cutters (Gauge 20 and Gauge 23), applying a fixed cut rate of 800 /min (13.33 Hz) at identical exposure times. The LaserVit-tips showed successful vitreous body removal for all laser settings and exposure times (Mark I: 6.2 g/min, Mark II: 8.2 g/min at 1500 cuts/min and 3 mJ, Mark II: 10.1 g/min, Mark III: 3.6 g/min at 6000 cuts/min at 3 mJ). Similar tip-dimensions (Gauge 22laser and Gauge 23cutter) showed comparable removal rates of 3.6 g/minlaser and 1.3 g/mincutter with settings of 6000 cuts/min at 3 mJ (laser) and 800 cuts/min for the mechanical cutter. A diode-pumped Q-switched Nd:YAG laser can successfully and gently remove vitreous body. The efficiency of the laser was comparable to that of mechanical cutters in terms of quantity of material removed per time unit.

Publisher

Springer Science and Business Media LLC

Subject

Multidisciplinary

Reference42 articles.

1. Blodi, C. F. David Kasner, MD the road to pars plana vitrectomy. Ophthalmol Eye Dis 14 8(Suppl 1), 1–4. https://doi.org/10.4137/OED.S40424 (2016).

2. Machemer, R., Buettner, H., Norton, E. W. & Parel, J. M. Vitrectomy: a pars plana approach. Trans. Am. Acad. Ophthalmol. Otolaryngol. 75(4), 813–820 (1971).

3. O’Malley, C. & Heintz, R. M Sr.. Vitrectomy with an alternative instrument system. Ann Ophthalmol 7(4), 585–588, 591–594 (1975).

4. Charles, S. The history of vitrectomy: innovation and evolution. Retina Today 9(10), 27–29 (2008).

5. de Oliveira, P. R. C., Berger, A. R. & Chow, D. R. Vitreoretinal instruments: vitrectomy cutters, endoillumination and wide-angle viewing systems. Int. J. Retina Vitreous 2, 28. https://doi.org/10.1186/s40942-016-0052-9 (2016).