Assessing the accuracy and reliability of application-based audiometry for hearing evaluation

Author:

Lee Seung Yeol,Seo Hee Won,Jung Seon Min,Lee Seung Hwan,Chung Jae Ho

Abstract

AbstractPure-tone audiometry (PTA) is the gold standard for assessing hearing loss. However, traditional PTA tests require specialized equipment, trained personnel, and a soundproof environment. Recently, smartphone-based PTA tests have been developed as an alternative method for hearing assessment. The aim of this study was to validate the accuracy and reliability of a smartphone application-based audiometry test. This study was conducted to assess the performance of application-based audiometry from November 2021 to January 2022. Pure-tone thresholds were measured using a smartphone application-based PTA test and compared with results obtained using a traditional audiometer in a sound-treated booth. The smartphone application used in this study was the "Care4Ear (Care4ear, version 1.0.6, MIJ Co., Ltd.)". Hearing thresholds less than 35 dB HL were classified as group A, 35-64 dB HL as group B, and 65 dB HL or greater as group C for the classification of hearing levels. We evaluated the accuracy of smartphone audiometry for each group and compared the results of frequency-specific hearing tests. Additionally, we examined the results of smartphone audiometry in individuals (n = 27) with asymmetric hearing loss. Seventy subjects completed both conventional audiometry and smartphone application-based hearing tests. Among the ears assessed, 55.7% were classified as group A, while 25.7% and 18.6% were classified as group B and group C, respectively. The average hearing threshold obtained from conventional pure-tone audiometry was 37.7 ± 25.2 dB HL, whereas the application-based hearing test yielded thresholds of 21.0 ± 23.0 dB HL. A significant correlation (r = 0.69, p < 0.01) was found between the average hearing thresholds obtained from the application-based and conventional pure-tone audiometry tests. The application-based test achieved a 97.4% hit rate for classifying hearing thresholds as class A, but lower rates of 22.2% for class B and 38.5% for class C. Notably, a discrepancy was observed between the hearing threshold measured by the application and the conventional audiometry for the worse ear with asymmetric hearing. The smartphone-based audiometry is a feasible method for hearing evaluation especially in persons with normal hearing. In cases of hearing loss or asymmetric hearing loss, the results of the application-based audiometry may be inaccurate, limiting its diagnostic utility.

Funder

Korea Health Industry Development Institute

Publisher

Springer Science and Business Media LLC

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3