Author:
Nanavaty Mayank A.,Naveed Hasan,Ashena Zahra,Mukhija Ritika
Abstract
AbstractThis study aimed to assess the surface quality of cap, stroma and lenticular surfaces created using low-energy femtosecond laser lenticule extraction (Ziemer FEMTO LDV Z8). Twenty-four porcine eyes were divided into four groups (n = 6 each): two with optimal laser power (32%) with posterior curvature equivalent to a spherical correction of -2D and -5D, respectively and the other two with high power (64%) with spherical correction of -2D and -5D respectively. Samples were analysed using scanning electron microscopy (SEM). Surface morphology was evaluated using a standard scoring system; surface relief, surface regularity, extent and position of irregularities were graded by four independent clinicians. Eyes with 32% power and -2D correction had significantly less size of the irregular area than those with -5D; however, no significant difference was found between the two groups with 64% power. When comparing eyes with -2D correction, the size of the irregular area was lesser with 32% power. Surface relief was lesser with -5D correction with 32% power than 64% power. Low-energy femtosecond laser lenticule extraction (Ziemer FEMTO LDV Z8) produces good surface quality results. There is a tendency for smoother surface stromal quality with lower power settings than a higher power.
Publisher
Springer Science and Business Media LLC