Author:
Urakawa Shinya,Hirashita Teijiro,Hirashita Yuka,Lowenfeld Lea,Gurram Krishna C.,Nishimura Makoto,Milsom Jeffrey W.
Abstract
AbstractEndoscopic submucosal dissection (ESD) is challenging in the right colon. Traction devices can make it technically easier. In this study, we evaluated a flexible grasper with articulating tip and elbow-like bending (IgE) through a double-balloon surgical platform (DESP), compared with an earlier generation grasper without elbow-like bending (Ig). The reach of Ig/IgE was investigated at eight locations using a synthetic colon within a 3D model. Using a fresh porcine colorectum, 4 cm pseudo-polyps were created at the posterior wall of the ascending colon. Fifty-four ESD procedures were performed using three techniques: standard ESD (STD), ESD using Ig (DESP + Ig), and ESD using IgE (DESP + IgE). IgE was able to reach the full circumference at all the locations, whereas the medial walls proximal to the descending colon were out of Ig’s reach. Compared with the STD, both DESP + Ig and DESP + IgE showed significantly shorter procedure time (STD vs. DESP + Ig vs. DESP + IgE = median 48.9 min vs. 38.6 vs. 29.9) and fewer injuries (1.5 vs. 0 vs. 0). Moreover, the DESP + IgE had a shorter procedure time than the DESP + Ig (p = 0.0025). The IgE with DESP increased instrument reach compared to Ig, and likely represented a traction tool for excision of large pseudo-polyps in the right colon.
Funder
This study was funded, in part, through a research grant from Lumendi LLC, Westport, CT, USA.
Publisher
Springer Science and Business Media LLC
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献