Biomechanical comparison of the femoral neck system and the dynamic hip screw in basicervical femoral neck fractures

Author:

Moon Jun-Ki,Lee Jung Il,Hwang Kyu-Tae,Yang Jae-Hyuk,Park Ye-Soo,Park Ki-Chul

Abstract

AbstractThe purpose of this study was to compare the fixation stability of proximal fragments and the mechanical characteristics in proximal femur models of basicervical femoral neck fracture fixed by the femoral neck system (FNS) versus the dynamic hip screw. The mean axial stiffness was 234 ± 35 N/mm in the FNS group and 253 ± 42 N/mm in the DHS group, showing no significant difference (p = 0.654). Mean values for x-axis rotation, y-axis rotation, and z-axis rotation after cycle load were 2.2 ± 0.5°, 6.5 ± 1.5°, and 2.5 ± 0.6°, respectively, in the FNS group and 2.5 ± 0.7°, 5.8 ± 2.1°, and 2.2 ± 0.9°, respectively, in the DHS group, showing no significant differences (p = 0.324, p = 0.245, and p = 0.312, respectively). The mean values of cranial and axial migration of screws within the femoral head were 1.5 ± 0.3 and 2.1 ± 0.2 mm, respectively, in the FNS group and 1.2 ± 0.3 and 2.4 ± 0.3 mm, respectively, in the DHS group, showing no significant differences (p = 0.425 and p = 0.625, respectively). The average failure load at vertical load was 1342 ± 201 N in the FNS group and 1450 ± 196 N in the DHS group, showing no significant difference (p = 0.452). FNS fixation might provide biomechanical stability comparable to that of DHS for treating displaced basicervical femoral neck fractures in young adults.

Publisher

Springer Science and Business Media LLC

Subject

Multidisciplinary

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3