Author:
Lee Yeoun Joo,Hwang Jae-Yeon,Ryu Hwaseong,Kim Tae Un,Kim Yong-Woo,Park Jae Hong,Choo Ki Seok,Nam Kyung Jin,Roh Jieun
Abstract
AbstractThis study assessed the image quality and diagnostic accuracy in determining disease activity of the terminal ileum of the reduced-dose computed tomography enterography using model-based iterative reconstruction in pediatric patients with Crohn’s disease (CD). Eighteen patients were prospectively enrolled and allocated to the standard-dose (SD) and reduced-dose (RD) computed tomography enterography (CTE) groups (n = 9 per group). Image quality, reader confidence in interpreting bowel findings, accuracy in determining active CD in the terminal ileum, and radiation dose were evaluated. Objective image quality did not show intergroup differences, except for image sharpness. Although reader confidence in detecting mural stratification, ulcer, and perienteric fat stranding of the RD-CTE were inferior to SD-CTE, RD-CTE correctly diagnosed active disease in all patients. The mean values of radiation dose metrics (SD-CTE vs. RD-CTE) were 4.3 versus 0.74 mGy, 6.1 versus 1.1 mGy, 211.9 versus 34.5 mGy∙cm, and 4.4 versus 0.7 mSv mGy∙cm for CTDIvol, size-specific dose estimation, dose-length product, and effective dose, respectively. RD-CTE showed comparable diagnostic accuracy to SD-CTE in determining active disease of the terminal ileum in pediatric CD patients. However, image quality and reader confidence in detecting ulcer and perienteric fat stranding was compromised.