Case report: Short-long-short mechanism triggering sustained ventricular tachycardia in a patient with a single-chamber ICD but inhibiting antitachycardia therapy

Author:

Schernthaner Christiana,Topf Albert,Motloch Lukas J.,Kraus Johannes,Hauptmann Laurenz,Hoppe Uta C.,Strohmer Bernhard

Abstract

IntroductionShort-long-short (SLS) sequences are an important cause of ICD pro-arrhythmia and can initiate both polymorphic and monomorphic ventricular tachycardias (VT). Depending on the programming of a single-chamber ICD, the interplay between SLS sequences and combined VT detection criteria can be responsible for withholding adequate anti-tachycardia pacing (ATP) or shock therapy.MethodsA 78-year-old patient with ICD was admitted to our emergency department after external cardioversion of a long-lasting VT with hemodynamic compromise. The interrogation of the ICD revealed an SLS sequence initiating a monomorphic VT at a rate of 171 bpm (350 ms). The VT discrimination of the implanted single-chamber ICD was based on the onset and stability criteria as the patient had a history of paroxysmal atrial fibrillation. The ICD was programmed that both criteria had to be met for VT detection and initiation of anti-tachycardia therapy.ResultsDue to the SLS sequence in combination with the programmed VT detection interval, the onset threshold was not fulfilled and inhibited adequate therapy. Some relatively slow VT beats following the SLS sequence resulted finally in a considerable delay in the declaration of the episode onset. As a first step, the threshold for VT detection was programmed to 150 instead of 160 bpm. To avoid SLS sequences and pause-dependent ventricular tachyarrhythmias, VVI backup stimulation was increased from 35 to 55 ppm. Besides, a device-specific algorithm called rate smoothing was activated as a potential preventive feature. On the 3-month follow-up, all sustained VT episodes were detected adequately by the reprogrammed device, resulting in appropriate anti-tachycardia pacing. After further refinement and less aggressive programming of rate smoothing, the patient remained free of symptoms and arrhythmias over a follow-up of more than 2.5 years, particularly since progression to permanent atrial fibrillation and pacing at a lower rate of 60 ppm.ConclusionsSLS sequences may initiate or trigger VT episodes. Misclassification of the true onset may occur in some ICD devices due to specific programming of VT detection criteria. If both criteria “Onset and Stability” have to be fulfilled, ICD therapy is not delivered despite ongoing VT. Anti-bradycardia backup pacing at a very low stimulation rate may facilitate SLS sequences in patients with ICD resembling a potential pro-arrhythmic mechanism. In case of gradual VT onset with some intervals slower than the programmed VT threshold, the detection rate has to be adjusted down to guarantee appropriate identification of the onset.

Publisher

Frontiers Media SA

Subject

Cardiology and Cardiovascular Medicine

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3