The application of fluoroscopic criteria to define leadless pacemakers implant position and the effect of location on device performance

Author:

ElRefai Mohamed1ORCID,Menexi Christina1,Abouelasaad Mohamed1,Roberts Paul12

Affiliation:

1. Cardiac Rhythm Management Research Department, University Hospital Southampton NHS Foundation Trust, Southampton, United Kingdom

2. Faculty of Medicine, University of Southampton, Southampton, United Kingdom

Abstract

Objective: Leadless pacemakers (LPs) were designed to avoid complications associated with transvenous pacing. To minimise risk of perforations, there is preference towards implanting LPs into the septum rather than the apex or free wall. An objective yet feasible way of characterising the LP location is currently lacking. We report a simple radiological method of defining LP position and our analysis of the impact of implantation site on performance of LPs. Methods: The first 100 LPs implanted at our UK centre were reviewed and the devices’ positions in fluoroscopy images and X-rays based on conventional criteria for lead positions and conventional practice for LPs positioning were assessed. The devices’ electrical parameters at implant and at the latest device follow-up were used to compare performance between implantation sites. Results: 35.6% of implants were in the apex. 31.1% in mid-septum, 16.7% in apical septum, 15.5% on the septal right ventricular inflow and 1.1% in the septal RV outflow tract. We had no major complications. Thresholds, R-wave amplitudes, and impedance averaged at 0.67 ± 0.41 V, 10.64 ± 5.30 mV, and 777.67 ± 201.67 Ohms, respectively, at the time of implantation, and 0.66 ± 0.39 V, 14.08 ± 6.14 mV, and 564.29 ± 96.76 Ohms at the last device check. There was no difference in the pacing thresholds or impedance between implant sites. Conclusions: We propose a simple, reproducible way of defining the LP location which can help standardise the assessment of the device location sites across LP implantation centres. Advances in knowledge: Emphasis on the safety and reliability of the leadless pacemakers in a real-world setting. Establishing the variation in the implantation sites for leadless pacemakers and reporting the effect of the implantation sites on the devices’ performance. We propose a simple, reproducible way of defining the LP location which can help standardise the assessment of the device location sites across LP implantation centres.

Publisher

Oxford University Press (OUP)

Subject

Radiology, Nuclear Medicine and imaging,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3