Quantitative assessment of variation in CBCT image technical parameters related to CBCT detector lateral-offset position

Author:

Santaella Gustavo Machado1ORCID,Rosalen Pedro Luiz2,Queiroz Polyane Mazucatto1,Haiter-Neto Francisco1,Wenzel Ann3,Spin-Neto Rubens3

Affiliation:

1. Department of Oral Diagnosis, Area of Oral Radiology, Piracicaba Dental School, University of Campinas, Piracicaba, Sao Paulo, Brazil

2. Department of Physiological Sciences, Area of Pharmacology, Piracicaba Dental School, University of Campinas, Piracicaba, Sao Paulo, Brazil

3. Department of Dentistry and Oral Health, Section of Oral Radiology, Aarhus University, Denmark, Europe

Abstract

Objectives: To assess the effect of CBCT detector position (aligned/lateral-offset) on image technical parameters (mean voxel value - MVV, standard deviation of voxel value (SDVV) distribution), comparing peripheral regions of interest (ROIs) to the central ROI in CBCT volumes. Methods: 40 CBCT volumes of a wax phantom were acquired in six units with aligned and/or lateral-offset detectors: Cranex 3Dx (CRA), Ortophos SL (ORT), Picasso Trio (PIC), Promax 3D Mid (PRO), Scanora 3D (SCA), and X1. Four image-acquisition protocols used an aligned detector, and four a lateral-offset detector. In each volume, 13 ROIs (12 peripheral and 1 central) were evaluated. MVV and SDVV of the peripheral ROI were compared to those of the central ROI in the volume. MVVD (the difference in percentage, between the MVV of a peripheral and the central ROI) was calculated. Results: For aligned-detectors, MVV increased (ORT and PRO) or decreased (CRA and X1) in the ROIs farther from the centre. For lateral-offset detectors, ROIs farther from the centre showed increased MVV. SDVV for most aligned detectors was lower, the nearer the ROI was to the centre. For lateral-offset detectors, it was lower for the peripheral ROIs, except with PIC. Range for MVVD was −32.8% to 22.8% for units with aligned detectors, and −20.7% to 69.5% for lateral-offset detectors. Conclusion: Lateral-offset detectors to acquire CBCT images significantly change SDVV distribution within the field-of-view, and lead to MVVD with increased range, compared to aligned detectors. This must be taken in consideration in the clinic, if voxel-value dependent measurements are to be performed.

Publisher

British Institute of Radiology

Subject

General Dentistry,Radiology, Nuclear Medicine and imaging,General Medicine,Otorhinolaryngology

Cited by 4 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3