Author:
Martinez Benjamin,Miller Thomas W.,Yalin Azer P.
Abstract
We present the development, integration, and testing of an open-path cavity ring-down spectroscopy (CRDS) methane sensor for deployment on small unmanned aerial systems (sUAS). The open-path configuration used here (without pump or flow-cell) enables a low mass (4 kg) and low power (12 W) instrument that can be readily integrated to sUAS, defined here as having all-up mass of <25 kg. The instrument uses a compact telecom style laser at 1651 nm (near-infrared) and a linear 2-mirror high-finesse cavity. We show test results of flying the sensor on a DJI Matrice 600 hexacopter sUAS. The high sensitivity of the CRDS method allows sensitive methane detection with a precision of ~10–30 ppb demonstrated for actual flight conditions. A controlled release setup, where known mass flows are delivered, was used to simulate point-source methane emissions. Examples of methane plume detection from flight tests suggest that isolated plumes from sources with a mass flow as low as ~0.005 g/s can be detected. The sUAS sensor should have utility for emissions monitoring and quantification from natural gas infrastructure. To the best of our knowledge, it is also the first CRDS sensor directly deployed onboard an sUAS.
Funder
U.S. Department of Agriculture
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Cited by
32 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献