Abstract
In Quartz-Enhanced PhotoAcoustic Spectroscopy (QEPAS) gas sensors, the acoustic wave is detected by the piezoelectric Quartz Tuning Fork (QTF). Due to its high-quality factor, the QTF can detect very low-pressure variations, but its resonance can also be affected by the environmental variations (temperature, humidity, …), which causes an unwanted signal drift. Recently, we presented the RT-QEPAS technique that consistently corrects the signal drift by continuously measuring the QTF resonance. In this article, we present an improvement of RT-QEPAS to fasten the QTF characterization time by adding a passive electronic circuit, which causes the damping of the QTF resonance. The damping circuit is optimized analytically and through SPICE simulation. The results are supported by experimental observations, showing a 70 times improvement of the relaxation times compared to the lone QTF, which opens the way to a fast and drift-free QEPAS sensor.
Funder
Agence Nationale de la Recherche
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Cited by
2 articles.
订阅此论文施引文献