Comparative Performance Assessment of Novel Fluorescence Immunoassay POCTs for Measuring Circulating Levels of Vitamin-D

Author:

Palermiti Alice1ORCID,Manca Alessandra1ORCID,Mastrantonio Fabrizio2,Maiese Domenico3,Curatolo Aurora3,Antonucci Miriam1,Simiele Marco3,De Nicolò Amedeo1ORCID,D’Avolio Antonio1ORCID

Affiliation:

1. Laboratory of Clinical Pharmacology and Pharmacogenetics, Department of Medical Sciences, University of Turin, 10149 Turin, Italy

2. A. Menarini Diagnostics, 50131 Florence, Italy

3. CoQua Lab s.r.l., 10149 Turin, Italy

Abstract

Vitamin D (Vit D) is a fat-soluble molecule acting like a hormone, and it is involved in several biological mechanisms such as gene expression, calcium homeostasis, bone metabolism, immune modulation, viral protection, and neuromuscular functions. Vit D deficiency can lead to chronic hypocalcemia, hyperparathyroidism, and many other pathological conditions; in this context, low and very low levels of 25-hydroxy-vitamin D (25-OH-D) were found to be associated with an increased risk of COVID-19 infection and the likelihood of many severe diseases. For all these reasons, it is important to quantify and monitor 25-OH-D levels to ensure that the serum/blood concentrations are not clinically suboptimal. Serum concentration of 25-OH-D is currently the main indicator of Vit D status, and it is currently performed by different assays, but the most common quantitation techniques involve immunometric methods or chromatography. Nevertheless, other quantitation techniques and instruments are now emerging, such as AFIAS-1® and AFIAS-10® (Boditech and Menarini) based on the immunofluorescence analyzer, that guarantee an automated system with cartridges able to give quick and reliable results as a point-of-care test (POCT). This work aims to compare AFIAS-1® and AFIAS-10® (Boditech and Menarini) Vit D quantitation with Ultra High-Performance Liquid Chromatography coupled with tandem mass spectrometry that currently represents the gold standard technique for Vit D quantitation. The analyses were performed in parallel on 56 samples and in different conditions (from fresh and frozen plasma) to assess the reliability of the results. Any statistically significant differences in methods, the fixed error, and the error proportional to concentration were reported. Results obtained in all conditions showed a good correlation between both AFIAS® instruments and LC-MS/MS, and we can affirm that AFIAS-1® and AFIAS-10® are reliable instruments for measuring 25-OH-D with accuracy and in a fast manner.

Publisher

MDPI AG

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3