Abstract
A gas chromatography-mass spectrometry (GC/MS) method for the determination of hexabromocyclododecane (HBCD) in expanded polystyrene and extruded polystyrene foam (EPS/XPS) was developed. The EPS/XPS samples were ultrasonically extracted with acetone and the extracts were purified by filtration through a microporous membrane (0.22 μm) and solid-phase extraction. The samples were analyzed using a GC/MS using the selected ion monitoring mode. The ions 157, 319 and 401 were selected as the qualitative ions, while ion 239 was chosen as the quantitative ion. An HBCD standard working solution with a concentration range of 1.0–50.0 mg/L showed good linearity. The detection limit of HBCD was 0.5 mg/kg, meeting the LPC limit (<100 or 1000 mg/kg). Six laboratories were selected to verify the accuracy of the method, and 10 samples were tested. The interlaboratory relative standard deviation range was 3.68–9.80%. This method could play an important role in controlling HBCD contamination in EPS/XPS.
Funder
National Key Research and Development Program of China
National Natural Science Foundation of China
Subject
Chemistry (miscellaneous),Analytical Chemistry,Organic Chemistry,Physical and Theoretical Chemistry,Molecular Medicine,Drug Discovery,Pharmaceutical Science
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
1. Hexabromocyclododecane;Encyclopedia of Toxicology;2024