Affiliation:
1. Department of Physics, College of Science, University of Hafr Al Batin, Al Jamiah, Hafr Al Batin 39524, Saudi Arabia
Abstract
In arid regions, the tritium concentration in groundwater is typically very low and often falls below the minimum detectable activity (MDA) of the conventional liquid scintillation counter (LSC). Therefore, to measure the tritium activity concentration, it is necessary to lower the detection limit so that the scintillation counter can detect it. In the present study, several methods are discussed which are effective at lowering the detectable activity of tritium. One of these methods is to enrich the tritium activity concentration by ten- to fortyfold of the initial concentration of the tritium. Twelve spiked samples with known amounts of tritium, five with high concentrations and seven with low concentrations, were enriched by the electrolysis process. The results indicated that enriching the tritium levels in groundwater lowers the MDA value. Other methods are minimizing background radiation using low-background materials for sample containers, increasing the measurement efficiency of the scintillation counter and counting time, and shielding the sample from environmental radiation using the shutter option in LSC. Moreover, reducing the number of interfering contaminants in the sample can lower the uncertainty in measuring the tritium concentration in the water sample, which is beneficial for detecting low-level tritium in water to ensure public health and safety.
Funder
Institutional Funding Project
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献