The Impact of Al2O3 Particles from Grit-Blasted Ti6Al7Nb (Alloy) Implant Surfaces on Biocompatibility, Aseptic Loosening, and Infection

Author:

Kocjančič Boštjan12,Avsec Klemen12,Šetina Batič Barbara3ORCID,Feizpour Darja3ORCID,Godec Matjaž3ORCID,Kralj-Iglič Veronika4,Podlipec Rok56,Cör Andrej7,Debeljak Mojca8,Grant John T.9ORCID,Jenko Monika310,Dolinar Drago1210

Affiliation:

1. Department for Orthopaedic Surgery, UMC Ljubljana, Zaloška 9, 1000 Ljubljana, Slovenia

2. Chair of Orthopedics, Faculty of Medicine, University of Ljubljana, Vrazov trg 2, 1000 Ljubljana, Slovenia

3. Institute of Metals and Technology, Lepi pot 11, 1000 Ljubljana, Slovenia

4. University of Ljubljana, Faculty of Health Sciences, Laboratory of Clinical Biophysics, 1000 Ljubljana, Slovenia

5. Laboratory for Biophysics, Jozef Stefan Institute, Jamova 39, 1000 Ljubljana, Slovenia

6. Ion Beam Center, Helmholtz-Zentrum Dresden-Rossendorf e.V., 01328 Dresden, Germany

7. Orthopaedic Hospital Valdoltra, Jadranska cesta 31, 6280 Ankaran, Slovenia

8. University Rehabilitation Institute Republic of Slovenia Soča, Linhartova 51, 1000 Ljubljana, Slovenia

9. Research Institute, University of Dayton, Dayton, OH 45469, USA

10. MD-RI Institute for Materials Research in Medicine, Bohoričeva 5a, 1000 Ljubljana, Slovenia

Abstract

For the improvement of surface roughness, titanium joint arthroplasty (TJA) components are grit-blasted with Al2O3 (corundum) particles during manufacturing. There is an acute concern, particularly with uncemented implants, about polymeric, metallic, and corundum debris generation and accumulation in TJA, and its association with osteolysis and implant loosening. The surface morphology, chemistry, phase analysis, and surface chemistry of retrieved and new Al2O3 grit-blasted titanium alloy were determined with scanning electron microscopy (SEM), X-ray energy-dispersive spectroscopy (EDS), transmission electron microscopy (TEM), X-ray photoelectron spectroscopy (XPS) and confocal laser fluorescence microscopy, respectively. Peri-prosthetic soft tissue was studied with histopathology. Blasted retrieved and new stems were exposed to human mesenchymal stromal stem cells (BMSCs) for 7 days to test biocompatibility and cytotoxicity. We found metallic particles in the peri-prosthetic soft tissue. Ti6Al7Nb with the residual Al2O3 particles exhibited a low cytotoxic effect while polished titanium and ceramic disks exhibited no cytotoxic effect. None of the tested materials caused cell death or even a zone of inhibition. Our results indicate a possible biological effect of the blasting debris; however, we found no significant toxicity with these materials. Further studies on the optimal size and properties of the blasting particles are indicated for minimizing their adverse biological effects.

Funder

Slovenian Research Agency, ARRS

Publisher

MDPI AG

Subject

General Materials Science

Reference54 articles.

1. Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030;Kurtz;J. Bone Joint Surg. Am.,2007

2. Surface chemistry and microstructure of metallic biomaterials for hip and knee endoprostheses;Jenko;Appl. Surf. Sci.,2018

3. Bobyn, J.D., Coiller, J.P., Mayor, M.B., McTighe, T., Tanzer, M., and Vaughn, B.K. (1993, January 5–7). Particulate debris in total hip arthroplasty: Problems and Solutions. Proceedings of the Scientific Exhibits at the 1993 AAOS Meeting, San Francisco, CA, USA.

4. Merola, M., and Affatato, S. (2019). Review Materials for Hip Prostheses: A Review of Wear and Loading Considerations. Materials, 12.

5. Delgado-Ruiz, R., and Georgios Romanos, G. (2018). Potential Causes of Titanium Particle, and Ion Release in Implant Dentistry: A Systematic Review. Int. J. Mol. Sci., 19.

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献