Abstract

Background and Objectives: In this study, we aimed to compare the physical properties of hole-implantable collamer lenses (H-ICLs) and implantable phakic contact lenses (IPCLs) and investigate their flexural and cell adhesion characteristics. Materials and Methods: Transverse compression load to achieve lens flexion and static Young’s modulus were measured in H-ICLs and IPCLs using designated equipment. Load was measured both with and without restraining the optic section of the lenses. Adhesion of iHLEC-NY2 cells to the lens surfaces was examined using phase-contrast microscopy, and cell proliferation activity was evaluated using WST-8 assay. Results: The H-ICL showed a greater tendency for transverse compression load compared to IPCL, while the IPCL showed a higher Young’s modulus with respect to the force exerted on the center of the anterior surface of the optic section. The joint between the optic section and haptic support in the IPCL was found to mitigate the effects of transverse compression load. Both lens types showed minimal cell adhesion. Conclusions: Our findings indicate that H-ICLs and IPCLs exhibit distinct physical properties and adhesive characteristics. The IPCL demonstrated higher Young’s modulus and unique structural features, while the H-ICL required greater transverse compression load to achieve the flexion required to tuck the haptic supports into place behind the iris to fix the lens. The observed cell non-adhesive properties for both lens types are promising in terms of reducing complications related to cell adhesion. However, further investigation and long-term observation of IPCL are warranted to assess its stability and potential impact on the iris. These findings contribute to a better understanding of the performance and potential applications of H-ICLs and IPCLs in ophthalmology.

Publisher

MDPI AG

Subject

General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3