Author:
Bocos-Bintintan Victor,Ratiu Ileana Andreea
Abstract
Sensitive real-time detection of vapors produced by toxic industrial chemicals (TICs) always represents a stringent priority. Hydrogen cyanide (HCN) is definitely a TIC, being widely used in various industries and as an insecticide; it is a reactive, very flammable, and highly toxic compound that affects the central nervous system, cardiovascular system, eyes, nose, throat, and also has systemic effects. Moreover, HCN is considered a blood chemical warfare agent. This study was focused toward quick detection and quantification of HCN in air using time-of-flight ion mobility spectrometry (ToF IMS). Results obtained clearly indicate that IMS can rapidly detect HCN at sub-ppmv levels in air. Ion mobility spectrometric response was obtained in the negative ion mode and presented one single distinct product ion, at reduced ion mobility K0 of 2.38 cm2 V−1 s−1. Our study demonstrated that by using a miniaturized commercial IMS system with nonradioactive ionization source model LCD-3.2E (Smiths Detection Ltd., London, UK), one can easily measure HCN at concentrations of 0.1 ppmv (0.11 mg m−3) in negative ion mode, which is far below the OSHA PEL-TWA value of 10 ppmv. Measurement range was from 0.1 to 10 ppmv and the estimated limit of detection LoD was ca. 20 ppbv (0.02 mg m−3).
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Cited by
9 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献