Drops of Capillary Blood Are Not Appropriate for Hemoglobin Measurement with Point-of-Care Devices: A Comparative Study Using Drop Capillary, Pooled Capillary, and Venous Blood Samples

Author:

De la Cruz-Góngora VanessaORCID,Méndez-Gómez-Humarán IgnacioORCID,Gaona-Pineda Elsa BereniceORCID,Shamah-Levy TeresaORCID,Dary Omar

Abstract

Population-based surveys matched by time but using different methodologies for determining hemoglobin (Hb) concentration have shown inconsistencies in estimating anemia prevalence. This study aimed to estimate measurement errors in Hb quantification in HemoCue 201+ using venous blood (VB) and capillary blood both drops (DCB) and pools (PCB), and compare the results against those of a reference method (VB analyzed in hematology analyzers based on the cyanmethemoglobin method). Children (n = 49), adult females (n = 50), and older adults (n = 50) were randomly allocated to donate VB (4 mL) and either DCB (three drops) or PCB (350 µL). Results in HemoCue were analyzed through Bland Altman and Lyn’s concordance against Hb concentration by the reference method. A positive average bias (systematic error) was found for the HemoCue (0.31 g/dL) using the same VB samples. This value was then subtracted from all readings carried out in the device. After this adjustment, DCB still produced a positive bias (0.42 ± 0.81 g/dL), and the variation of single results was ±1.6 g/dL (95% CI). PCB and VB performed similarly; the average bias was negligible (−0.02 ± 0.36 and 0.00 ± 0.33 g/dL, respectively) and the variation of the results (95% CI) was ±0.7 g/dL or lower. Lyn’s concordance values were 0.86, 0.96, and 0.98 for DCB, PCB, and VB, respectively. Random variation using DCB is too large to approximate the true Hb values, and therefore DCB should be discontinued for diagnosing anemia both in individuals and in populations.

Funder

Mexican Ministry of Health

Publisher

MDPI AG

Subject

Food Science,Nutrition and Dietetics

Reference27 articles.

1. World Health Organization (WHO) (2022, November 11). Haemoglobin Concentrations for the Diagnosis of Anaemia and Assessment of Severity. Vitamin and Mineral Nutrition Information System, Available online: http://Www.Who.Int/Vmnis/Indicators/Haemoglobin.Pdf.

2. World Health Organization (WHO) (2014). Global Nutrition Targets 2025: Anaemia Policy Brief (WHO/NMH/NHD/14.4).

3. (2022, November 11). HemoCue Hb 201 + Operating Manual Manuel d’utilisation, Available online: https://www.michigan.gov/documents/mdhhs/hemocue_operating_manual_app_d_hb201manual_543516_7.pdf.

4. (2022, November 11). HemoCue Hb 201+ System Specifications. Available online: https://medical.andonline.com/wp-content/uploads/2021/12/Hb-201-Brochure.pdf.

5. Factors Associated with Anaemia in a Nationally Representative Sample of Nonpregnant Women of Reproductive Age in Nepal;Ford;Matern. Child Nutr.,2022