Abstract
Radiation dose management of medical staff has become increasingly important. Particularly, based on the statement by the International Commission on Radiological Protection (ICRP) in 2011, a new lower equivalent dose limit for the eye lens is being established in each country. Although many reports have discussed the occupational radiation dose in interventional radiology (IR), few studies have examined the dose during computed tomography (CT) examinations. This study investigated the radiation dose exposure to medical staff present in the CT room during irradiation, with particular focus on the exposure to eye lens. The radiation dose exposure to those who assist patients during head, chest and upper abdomen CT examination was measured in a phantom study. The radiation dose exposure with scattered radiation was never negligible (i.e., high); Hp(3) was the highest in head CT examination, at 0.44 mSv per examination. Furthermore, the shielding effect of radiation protection glasses was large, and radiation protection glasses are useful tools for the medical staff who are involved in CT examinations. The justification and optimisation should be carefully considered in assistant actions.
Subject
Fluid Flow and Transfer Processes,Computer Science Applications,Process Chemistry and Technology,General Engineering,Instrumentation,General Materials Science
Cited by
6 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献